Single use interventional kit/tray that provides the two main components necessary to perform a "safe" MRI interventional procedure. An ideal base to build your "custom" tray around.
Quantity Description
1 #11 blade non-ferromagnetic scalpel with plastic handle
1 20-gauge non-ferromagnetic injection needle 9cm
1 CSR Outer Wrap 28 x 28
1 Prep Tray 10x5x2
*Non-Clinical testing demonstrated that this product is MR conditional and can be used in the MRI environment according to the following conditions:
1. Static magnetic of 3-Tesla or less
2. Highest spatial gradient magnetic field of 720-Gauss/cm or less
Important note: This product is intended for use inside of the MRI environment (in the MR system room). It should not be utilized directly inside of the MR system (inside the bore of the scanner) during its operation (scanning). As such, the assessment of magnetic field interactions for this product specifically involved evaluations of translational attraction and torque in relation to exposure to a 3-Tesla MR system. This product has not been tested for MRI-related heating and artifacts.